APPLICATIONS
- Urban waste water treatment
- Industrial effluent treatment
- Surface Water monitoring
- Sea Water
- Drinking Water
MOUNTING AT 4 ELECTRODES
The electrode works with a technology in 4 electrodes an alternating current of constant-voltage is established between a primary’s pair of electrodes in graphite. The secondary’s electrodes in platinum allow of regulate the voltage imposed to primary’s electrodes to reflect of the fouling The voltage measured between the primary’s electrodes is in function of the resistance of place and so of the conductivity.
DIGITAL TECHNOLOGY
The “Smart” Digital SE sensor Stores calibration and history data within the sensor. This allows you a “plug and play” system without re-calibration.
Measures
Measure principle | Conductivity sensor with 4 electrodes (2 graphic, 2 platinum) |
Measure ranges | 0-200,0 uS/cm 0-2000 uS/cm 0,00 – 20,00 mS/cm 0,0 – 200,0 mS/cm |
Resolution | 0,01 to 1 according the range |
Accuracy | ±1 % of the full range |
Measure range salinity | 5-60 g/kg |
Measure range TDS-KCI | 0-133 000 ppm |
Response time | < 5s |
Working temperature | 0 °C to + 50 °C |
Temperature compensation | NTC |
Stocking temperature | -10 °C to + 60 °C |
Sensor
Dimensions | Diameter : 27 mm ; Length : 157 mm |
Weight | 300 g(sensor +cable 3 meters) |
Material | PVC, DELRIN, stainless steel |
Maximum pressure | 5 bars |
Connection | 9 armored connectors, polyurethane jacket, bare – wires or waterproof Fisher connector |
Degree of protection | IP68 |