CONDUCTIVITY / SALINITY

  • Home
  • CONDUCTIVITY / SALINITY
  • admin_sune
  • No Comment
  • 4 March, 2020

CONDUCTIVITY / SALINITY

APPLICATIONS

  • Urban waste water treatment
  • Industrial effluent treatment
  • Surface Water monitoring
  • Sea Water
  • Drinking Water

 

MOUNTING AT 4 ELECTRODES

The electrode works with a technology in 4 electrodes an alternating current of constant-voltage is established between a primary’s pair of electrodes in graphite. The secondary’s electrodes in platinum allow of regulate the voltage imposed to primary’s electrodes to reflect of the fouling The voltage measured between the primary’s electrodes is in function of the resistance of place and so of the conductivity.

DIGITAL TECHNOLOGY

The ‘Smart” Digital SE sensor Stores calibration and history data within the sensor. This allows you a “ plug and play system without re-calibration.

Measures

Measure principle  Conductivity sensor with 4 electrodes (2 graphic, 2 platinum)
Measure ranges 0-200,0 uS/cm

0-2000 uS/cm

0,00 – 20,00 mS/cm

0,0 – 200,0 mS/cm

Resolution 0,01 to 1 according the range
Accuracy ±1 % of the full range
Measure range salinity 5-60 g/kg
Measure range TDS-KCI 0-133 000 ppm
Response time < 5s
Working temperature 0 C to + 50 C
Temperature compensation NTC
Stocking temperature -10 C to + 60 C

 Sensor

Dimensions Diameter : 27 mm ; Length : 157 mm
Weight 300 g(sensor +cable 3meters)
Material PVC, DELRIN, stainless steel
Maximum pressure 5 bars
Connection 9 armored connectors, polyurethane jacket, bare – wires or waterproof Fisher connector
Degree of protection IP68

 

Download catalog